Base unit
Note | |||
Version HC | |||
Clean air conditioner grilles, pump room ventilation inlet, and sample fan outlet | 1x Month | Howto | check/brush over each visit |
Unmounting and cleaning the complete air conditioner | 1x year | Howto | depending on the air quality on site |
Version 2D, 4D, PM | |||
Clean air conditioner grilles | 1x Month | Howto | check/brush over each visit |
All Versions | |||
Sample Filter ø47mm | 1-4x month | Howto | depending on the air quality on site, it never should become completely black |
Exdented life Filter | 1-4x year | Howto | depending on the air quality on site, at least once a year |
Exchange zero air adsorbants and DFU filter | 4x Year | Howto | depending on the air quality on site |
Rebuild Pump (new K model) | 1x Year | Howto | |
Rebuild Pump (old model) | 1x Year | Howto | |
Sampling system & head cleaning | 1x Year | Howto |
NOx Module
All Versions | Note | ||
Zero/Span Calibration and Converter efficiency test | 1-4x year | Howto | depending on your country regulation, but at least once a year |
Version API Bench | |||
Clean reaction cell window | 1x year | Howto | |
Exchange sintered filters and o-rings | 1x year | Howto | |
Exchange 1 DFU filter | 1x year | Howto | |
Version Thermo Bench | |||
Clean reaction cell window | Howto | When sensitivity is decreasing | |
Exchange 1 DFU filter | 1x year | Howto |
SO2 Module
All Versions | Note | ||
Zero/Span Calibration | 1-4x year | Howto | depending on your country regulation, but at least once a year |
Version API Bench | |||
SOx Scrubber | 1x year | Howto | Only with option H2S |
Version Thermo Bench | |||
SOx Scrubber | 1x year | Howto | Only with option H2S |
CO Module
All Versions | Note | ||
Zero/Span Calibration | 1-4x year | Howto | depending on your country regulation, but at least once a year |
Version API benches | |||
Exchange 1 DFU filter | 1x year | Howto | |
Version Thermo benches | |||
Exchange 1 DFU filter | 1x year | Howto |
O3 Module
All Versions | Note | ||
Zero/Span Calibration | 1-4x year | Howto | depending on your country regulation, but at least once a year |
Version API benches | |||
Exchange ozone scrubber | typically every 4 years | Howto | depending on the air quality on site, change if sensitivity is decreasing |
Version Thermo benches | |||
Exchange ozone scrubber | typically every 4 years | Howto | depending on the air quality on site, change if sensitivity is decreasing |
BTEX Module
All Versions | Note | ||
Zero/Span Calibration | 1-4x year | Howto | depending on your country regulation, but at least once a year |
Exchange zero air filter | 1x year | Howto | |
Exchange PID Lamp | 1x year | Howto | change once a year or if sensitivity is decreasing |
multiPM Module
All Versions | Note | ||
Exchange DFU filter | 1x year | Howto | depending on the air quality on site, but at least once a year |
Sampling system & head cleaning | 1x year | Howto | depending on the air quality on site |